summaryrefslogtreecommitdiffstats
path: root/Documentation
diff options
context:
space:
mode:
authorGravatar Robert C. Helling <helling@atdotde.de>2015-01-30 11:57:57 +0100
committerGravatar Dirk Hohndel <dirk@hohndel.org>2015-01-30 07:18:06 -0800
commit3788168114f6d44f4a8ca50face08081d912e2e8 (patch)
treeb806af7e49cecb14216f66057ad9c7cd686f665b /Documentation
parente9558d848c6992848bbf10126cd1b05c80db8091 (diff)
downloadsubsurface-3788168114f6d44f4a8ca50face08081d912e2e8.tar.gz
User manual additions for rebreather modes
...in particular for planning rebreather dives. Signed-off-by: Robert C. Helling <helling@atdotde.de> Signed-off-by: Dirk Hohndel <dirk@hohndel.org>
Diffstat (limited to 'Documentation')
-rw-r--r--Documentation/user-manual.txt30
1 files changed, 27 insertions, 3 deletions
diff --git a/Documentation/user-manual.txt b/Documentation/user-manual.txt
index cb77a0586..773524600 100644
--- a/Documentation/user-manual.txt
+++ b/Documentation/user-manual.txt
@@ -1683,7 +1683,7 @@ CCR equipment provide actual measurements of pO~2~, derived from oxygen sensors.
In this case the graph for oxygen partial pressure
should be fairly flat, reflecting the setpoint settings during the dive. The mean
pO~2~ is NOT the mean oxygen partial pressure as given by the CCR equipment, but
-a value xalculated by _Subsurface_ as follows:
+a value calculated by _Subsurface_ as follows:
- For TWO O~2~ sensors the mean value of the two sensors are given.
@@ -1691,6 +1691,8 @@ a value xalculated by _Subsurface_ as follows:
differences of more than 0,1 bar in the simultaneous readings of different sensors are
treated as spurious. If one of the three sensors provides spurious data, it is ignored.
+- If no sensor data is available, the pO~2~ value is assumed to be equal to the setpoint.
+
The mean pO~2~ of the sensors is indicated with a green line,
The oxygen setpoint values as well as the readings from the individual
@@ -2629,10 +2631,13 @@ This panel allows two type of selections:
** _Default CCR setpoint for dive planning:_ Specify the O~2~ setpoint for a
CCR dive plan. This determines the pO~2~ that will be maintained
- during a particular dive.
+ during a particular dive. This is the setpoint that is used at the start
+ of any CCR dive. Setpoint changes during the dive can be added via the
+ profile context menu.
** _pSCR O~2~ metabolism rate:_ For a semiclosed rebreather (pSCR) system, this is the
- volume of oxygen used by a diver each minute. Set this value for pSCR dive planning.
+ volume of oxygen used by a diver each minute. Set this value for pSCR dive planning
+ and decompression calculations.
** _pSCR ratio:_ For pSCR equipment the dump ratio is the ratio of gas released to the
environment to that of the gas recirculated to the diver. Set this value for a
@@ -2857,6 +2862,25 @@ level is indicated in the _Dive Plan Details_. This duration INCLUDES the transi
time to get to that level. However, if the _Display transition in deco_ option is checked,
the transitions are shown separately from the segment durations at a particular level.
+=== Planning rebreather dives
+
+To plan a dive using a passive semi-closed rebreather (pSCR), select "pSCR" rather than
+"Open circuit" in the Planning tab. This takes the oxygen drop in the rebreather into account.
+The parameters of the pSCR can be set in the preferences in the Graph tab. A warning apears
+in the dive plan details, if the pO~2~ drops below what is considered a save value. For pSCR dives,
+the gas consumption calculation takes into account the pSCR ratio (default 10:1).
+
+To plan a dive using a closed circuit rebreather (CCR), select the "CCR" option in the PLannig tab.
+This makes all user entered segments use the default setpoint value (configured in the Preferences
+in the Graph tab). The gases from the gas list are considered to be diluent gases (the oxygen tank is
+implicitely assumed to be present).
+
+Then, individual setpoints can be set for dive segments in the "Dive planner points" table. A zero setpoint
+means the diver bails out to open circuit mode for that segment. The decompression is always calculated
+with the setpoint of the last manually entered segment. So, to plan an bail out ascent for a
+CCR dive, add a one minute dive segment in the end with a setpoint value of 0. The decompresseion
+algorithm does not switch decogases automatically in CCR mode (but of course for bail out ascents).
+
[[S_Replan]]
=== Modifying an existing dive plan